Özet
Purpose: To investigate the dosimetric feasibility of linear accelerator-based intensity-modulated total marrow irradiation (IM-TMI) in patients with hematologic malignancies. Methods and Materials: Linear accelerator-based IM-TMI treatment planning was performed for 9 patients using the Eclipse treatment planning system. The planning target volume (PTV) consisted of all the bones in the body from the head to the mid-femur, except for the forearms and hands. Organs at risk (OAR) to be spared included the lungs, heart, liver, kidneys, brain, eyes, oral cavity, and bowel and were contoured by a physician on the axial computed tomography images. The three-isocenter technique previously developed by our group was used for treatment planning. We developed and used a common dose-volume objective method to reduce the planning time and planner subjectivity in the treatment planning process. Results: A 95% PTV coverage with the 99% of the prescribed dose of 12 Gy was achieved for all nine patients. The average dose reduction in OAR ranged from 19% for the lungs to 68% for the lenses. The common dose-volume objective method decreased the planning time by an average of 35% and reduced the inter- and intra- planner subjectivity. Conclusion: The results from the present study suggest that the linear accelerator-based IM-TMI technique is clinically feasible. We have demonstrated that linear accelerator-based IM-TMI plans with good PTV coverage and improved OAR sparing can be obtained within a clinically reasonable time using the common dose-volume objective method proposed in the present study.
| Orijinal dil | İngilizce |
|---|---|
| Sayfa (başlangıç-bitiş) | 1256-1265 |
| Sayfa sayısı | 10 |
| Dergi | International Journal of Radiation Oncology Biology Physics |
| Hacim | 79 |
| Basın numarası | 4 |
| DOI'lar | |
| Yayın durumu | Yayınlandı - 15 Mar 2011 |
| Harici olarak yayınlandı | Evet |
BM SKH
Bu sonuç, aşağıdaki Sürdürülebilir Kalkınma Hedefine/Hedeflerine katkıda bulunur
-
SKH 3 Sağlık ve Kaliteli Yaşam
Parmak izi
Linear accelerator-based intensity-modulated total marrow irradiation technique for treatment of hematologic malignancies: A dosimetric feasibility study' araştırma başlıklarına git. Birlikte benzersiz bir parmak izi oluştururlar.Bundan alıntı yap
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver