Abstract
Introduction: Chromatographic methods for determination of betamethasone or dexamethasone in their pharmaceutical preparations by Ultra Performance Liquid Chromatography-Tandem Mass Spectrometry were developed. Method: Analytes were eluted from Acquity BEH C18 (50 mm × 2.1 mm, 1.7 μm) column using stepwise gradient with mobile phase consisted of acetonitrile and 0.1% (v/v) acetic acid in water. Quantitation of dexamethasone and betamethasone was performed using 451>361 m/z on MRM mode. Results: Methods were linear over the concentration range from 10 to 1500 ng mL-1for both analyte. Limits of detection for BTM or DXM were 1 ng mL-1The highest relative error and relative standard deviation values were not over 2%. Conclusion: Presented methods were specific, rugged and robust, successfully applied to the analysis of their pharmaceutical preparations.
| Original language | English |
|---|---|
| Pages (from-to) | 68-75 |
| Number of pages | 8 |
| Journal | Current Pharmaceutical Analysis |
| Volume | 14 |
| Issue number | 1 |
| DOIs | |
| Publication status | Published - 1 Jan 2018 |
Keywords
- Ampoule
- Betamethasone
- Dexamethasone
- Drops
- Tablet
- UPLC-MS/MS
Fingerprint