Skip to main navigation Skip to search Skip to main content

Cloud point extraction for the determination of trace copper in water samples by flame atomic absorption spectrometry

Research output: Contribution to journalArticlepeer-review

54 Citations (Scopus)

Abstract

Cloud point extraction has been used for preconcentration of trace amount of copper. The analyte was complexed with 1,2-dihydroxyanthraquinone-3-sulfonic acid sodium salt, and octylphenoxypolyethoxyethanol was added as a surfactant. The effects of experimental conditions such as pH, concentration of chelating agent and surfactant, equilibration temperature and time on cloud point extraction were studied. Under the optimum conditions, preconcentration of a 12 mL sample solution permitted the detection of 1.07 ng mL-1 of copper with an enrichment factor of 21. The method was applied to the determination of trace copper in water samples.

Original languageEnglish
Pages (from-to)107-112
Number of pages6
JournalMicrochimica Acta
Volume162
Issue number1-2
DOIs
Publication statusPublished - Jul 2008

Keywords

  • ARS
  • Cloud point extraction
  • Copper
  • FAAS
  • Preconcentration

Fingerprint

Dive into the research topics of 'Cloud point extraction for the determination of trace copper in water samples by flame atomic absorption spectrometry'. Together they form a unique fingerprint.